                                advertising feature
                                                                                                                                                                                                                                                                                                                          application notes
                                                                                                                                                                  ™
                                The Simple Western : a gel-free, blot-free, hands-free 
                                Western blotting reinvention
                                Western blotting is considered the gold standard for protein detection and characterization. Although 
                                improvements to individual aspects of Western methodologies have been developed in recent years, 
                                                                                                                                                                                                                                                                                                                    ™ 
                                none has integrated the entire process onto a single platform. ProteinSimple has developed Simple 
                                                              ™
                                Western  assays for protein sizing and quantitative immunodetection as an alternative to traditional 
                 ed.ed.                                                                                                                                                                                                           ™
                 vv             Western blot analysis. Assays are performed on Simon , an instrument that integrates and automates 
                 esereser       all manual operations associated with Western blotting.
                 ights rights r
                 All rAll r     Introduction                                                                                                                                                                                   The separated proteins are immobilized to the capillary wall via a 
                     
                                Western blotting is the most widely used and accepted methodology                                                                                                                              proprietary, photoactivated capture chemistry. Target proteins are then 
                  Inc. Inc.     for protein detection and was first reported in the literature over 30                                                                                                                         identified with a primary antibody and subsequent immunodetection 
                 ica,ica,                                 1,2                                                                                                                                                                  using a horseradish peroxidase (HRP)-conjugated secondary antibody 
                                years ago                        . Although Western blotting is a proven technique, it 
                                is plagued by poor reproducibility, lack of accurate quantitation,                                                                                                                             and chemiluminescent substrate. Molecular weight and signal for 
                 AmerAmer       extensive time to result and reliability issues. Improvements have been                                                                                                                        immunodetected proteins are automatically reported. Simultaneous 
                 e e 
                 urur           made to reagents and individual steps within the Western blotting                                                                                                                              analysis of up to 12 samples can be performed in a single experiment, 
                  Nat Nat       process over the years, but none has fully overcome the challenges and                                                                                                                         and results are available in 3–5 hours. The software reports molecular 
                 11             bottlenecks still experienced by researchers today. The Simple Western                                                                                                                         weight, area, percent area and signal to noise for each protein detected.
                 11
                                is a reinvention of the entire Western blot, automating all steps from 
                 © 20© 20       protein loading and separation, immunoprobing, washing, detection                                                                                                                                    a                                                                                  b
                                and quantitative analysis of data, finally giving researchers a complete, 
                                                                                                                                                                                                                                                                                                                                                                      MCF-7          MCF-7MCF-7 Ladder
                                                                                                                                                                                                                                                                                                                                        MCF-7MCF-7     MCF-7MCF-7MCF-7     MCF-7MCF-7
                                walk-away solution. Manual factors that can negatively impact                                                                                                                                                                                                                                                      MCF-7
                                                                                                                                                                                                                                                                                                                                                                                                   250 kDa
                                reproducibility, quantitation, time to result and overall reliability of                                                                                                                                                                                                                                                                                           150 kDa
                                the generated data are eliminated.                                                                                                                                                                                                                                                                                                                                 100 kDa
                                                                                                                                                                                                                                                                                                                                                                                                   75 kDa
                                                                                                                                                                                                                                                                                                                                                                                                   50 kDa
                                Simple Western basics                                                                                                                                                                                                                                                                                                                                              37 kDa
                                                                                                                                                                                         3                                                                                                                                                                                                         25 kDa
                                Samples are prepared following conventional procedures .Samples                                                                                                                                                                                                                                                                                                    20 kDa
                                are then mixed with Simple Western Sample Buffer and standards to a                                                                                                                                                                                                                                                                                                15 kDa
                                final concentration of 1 μg/μL, reduced and denatured. The prepared                                                                                                                                                                                                                                                                                                10 kDa
                                samples, primary and secondary antibodies and chemiluminescent 
                                substrate are dispensed in microliter volumes into designated wells in a 
                                low-volume 384-well assay plate. Simple Western assay buffers, nano-
                                volume capillaries and the prepared assay plate are placed in Simon, 
                                which carries out all assay steps automatically. Proteins are separated                                                                                                                             Figure 1 | Comparison of phosphoinositide 3-kinase detection via Simple 
                                                                                                                                                                                                                                    Western and Western blot. Ten replicates (n = 10) of MCF-7 lysate (Santa Cruz 
                                in capillaries as they migrate through a stacking and separation matrix.                                                                                                                            Biotechnology p/n SC-2206) at 1 μg/μL were analyzed using a PI3K-specific 
                                                                                                                                                                                                                                    primary antibody (Cell Signaling p/n 3011). The predicted molecular weight 
                                                                                                                                                                                                                                    of PI3K is 110 kDa. (a) Simple Western lane view with results automatically 
                                Uyen Nguyen, Nick Squaglia, Annegret Boge & Peter A Fung                                                                                                                                            analyzed by system software. (b) Western blot image captured on FluorChem 
                                                                                                                                                                                                                                    M with results analyzed manually in AlphaView software (both ProteinSimple). 
                                ProteinSimple, Santa Clara, California, USA. Correspondence should be addressed to P.A.F.                                                                                                           Replicate data averages are summarized in Table 1.
                                (peter.fung@proteinsimple.com).
                                                                                                                                                                                                                                                                                                                   nature methods | NOVEMBER 2011 | v
                                                                                                                                                                                                                                                                                                                                                                                                        advertising feature
                                         application notes
                                      table 1 | Summarized results for the Simple Western and Western blot data shown in Figure 1.
                                                                 pi3K                                                             mW (kda)                                                                 % cv (mW)                                                                        signal                                                       % cv (signal)                                                              signal to noise
                                      Simple Western                                                                                       107                                                                         0.5                                                                  33747                                                                        8.7                                                                           66
                                      Western blot                                                                                         114                                                                         2.2                                                                212295                                                                         8.7                                                                          9.3
                                      PI3K, phosphoinositide 3-kinase. MW, molecular weight. CV, coefficient of variation.
                                      More quantitative and reproducible results                                                                                                                                                                                          Wider dynamic range
                                      Reproducibility of results from a traditional Western blot is a common                                                                                                                                                              Simple Western assays have a linear dynamic range of approximately 
                                      challenge for researchers due to lack of standardized procedures and                                                                                                                                                                three orders of magnitude depending upon the protein target. As 
                                      the multiple handling steps that introduce experimental variability.                                                                                                                                                                shown in Figure 2a, the dynamic range for glycogen synthase kinase-
                                                                                                                                                                                                                                                                                                                                                                                                                                              2
                                      Because the Simple Western assay is fully automated, results are                                                                                                                                                                    3α (GSK-3α) in a K562 lysate was 3.3 logs with an R  value of 0.999. 
                                      more reproducible than those generated via Western blot. Overall                                                                                                                                                                    For Western blot analysis on the same lysate samples using the same 
                                      quantitation is vastly improved as blot transfer is not required,                                                                                                                                                                   antibody (Fig. 2b), a less linear response was observed, with a dynamic 
                                      thus eliminating any inconsistencies in protein transfer. Figure 1                                                                                                                                                                                                                                                   2
                                                                                                                                                                                                                                                                          range of 2.5 logs and an R  value of 0.971.
                                      demonstrates the reproducibility and accuracy of a Simple Western 
                                      assay compared to Western blot for detection of phosphoinositide                                                                                                                                                                    Summary
                    ed.ed.            3-kinase (PI3K) expression in an MCF-7 lysate. Simple Western assay                                                                                                                                                                 The Simple Western is the first fully automated and complete solution 
                    vv                data (Fig. 1a) is represented by a software-generated lane view image,                                                                                                                                                              for protein detection and characterization, representing a true 
                    esereser          and protein size, signal intensity and area of the chemiluminescent                                                                                                                                                                 reinvention of Western blotting. Researchers are now able to simply 
                                      signal are reported. Western blot data (Fig. 1b) was generated following                                                                                                                                                            load their samples, press start, walk away and return a few hours later 
                    ights rights r    a standard protocol, and the fluorescent image was captured using a                                                                                                                                                                 to fully analyzed experimental results. Simon automates the entire 
                                      traditional imager and analysis software. Results are summarized in                                                                                                                                                                 process from start to finish and eliminates all hands-on labor, offering 
                    All rAll r        Table 1.                                                                                                                                                                                                                            significant time savings and drastically decreasing time to result. The 
                        
                                                                                                                                                                                                                                                                          high quality of data generated is considerably more reproducible 
                     Inc. Inc.                                                                                                                                                                                                                                            between users and over time. In addition, the process variability, blot 
                    ica,ica,                  a                                                                                                                                                                                                                           transfer and manual analysis that made traditional Western blot results 
                                                                                                                                                                                                                                                                          semi-quantitative at best are eliminated, allowing highly quantitative 
                    AmerAmer                                                                                                                                                                                                                                              results to be obtained over a wide dynamic range. Up to 12 samples 
                    e e                                                                                                                                                Dynamic range for GSK-3α                                          1,000,000
                    urur                                                                                                                                                      (Simple Western)                                                                            can be analyzed in 3–5 hours, and targets between 15–150 kDa 
                                                                                                                                                                                                                                         100,000                          can be detected. Simple Western assays run on Simon also facilitate 
                     Nat Nat
                    11                                                                                                                                                                                                                   10,000                           standardization of laboratory processes, and provide data in a format 
                    11
                                                                                                                                                                                                                                         1,000                            that can be easily shared between multiple users and facilities. For 
                    © 20© 20                                                                                                                                                                                         R = 0.9995
                                                                                                                                                                                                                                         100                              more information please visit proteinsimple.com
                                                                                                                                                                                                                                         10
                                                                                                                                                   0.0001           0.001            0.01             0.1              1               10                                 1.           Towbin, H. et al. Electrophoretic transfer of proteins from polyacrylamide gels 
                                                                                                                                                                                   K562 lysate [mg/mL]                                                                                 to nitrocellulose sheets: procedure and some applications. Proc. Natl. Acad. Sci. 
                                                                                                                                                                                                                                                                                       USA 76, 4350–4354 (1979).
                                              b                                                                                                                                                                                                                           2.           Renart, J. et al. Transfer of proteins from gels to diazobenzyloxymethyl-paper 
                                                                                                                                                                                                                                                                                       and detection with antisera: a method for studying antibody specificity and 
                                                                                                                                                                                                                                                                                       antigen structure. Proc. Natl. Acad. Sci. USA 76, 3116–3120 (1979).
                                                               3 mg/ml                                 0.001 mg/ml
                                                                              0.11mg/ml0.03 mg/ml0.003 mg/ml
                                                                         0.33 mg/ml0.03 mg/ml0.01 mg/ml          0.0001 mg/ml
                                                                    1 mg/ml                                 0.0003 mg/ml                                                                                                                                                  3.           Gallager, S.R. One-dimensional SDS gel electrophoresis of proteins, basic 
                                                                                                                              250 kDa                                Dynamic range for GSK-3α                                            1,000,000                                     protocol 1. In Current Protocols in Protein Science (eds. Coligan, J.E. et al.) 
                                                                                                                              150 kDa                                          (Western blot)
                                                                                                                                                                                                                                         100,000                                       10.1.1 -10.1.34 (John Wiley & Sons, Somerset, N.J., 1995). 
                                                                                                                              100 kDa
                                                                                                                              75 kDa
                                                                                                                                                                                                                                         10,000
                                                                                                                              50 kDa                                                                             R = 0.9707
                                                                                                                               37 kDa                                                                                                    1,000
                                                                                                                               25 kDa                                                                                                    100
                                                                                                                               20 kDa
                                                                                                                              15 kDa                                                                                                     10
                                                                                                                              10 kDa                 0.0001          0.001           0.01             0.1              1             10
                                                                                                                                                                                K562 lysate [mg/mL]
                                           Figure 2 | Comparison of Simple Western and Western blot dynamic range. 
                                           K562 cells lysed in Bicine/CHAPs buffer were serially diluted and analyzed 
                                           using a glycogen synthase kinase-3α (GSK-3α) antibody (Cell Signaling p/n 
                                           4818). (a) Simple Western lane view with quantitative results automatically 
                                           generated in system software. (b) Western blot results captured using 
                                           FluorChem M with quantitation manually performed using AlphaView                                                                                                                                                               This article was submitted to Nature Methods by a commercial organization 
                                           software (both ProteinSimple). Coefficient plots were generated in                                                                                                                                                             and has not been peer reviewed. Nature Methods takes no responsibility for 
                                                                    ® Excel® for both methods.
                                           Microsoft                                                                                                                                                                                                                      the accuracy or otherwise of the information provided.
                                      vi | NOVEMBER 2011 | nature methods
The words contained in this file might help you see if this file matches what you are looking for:

...Advertising feature application notes the simple western a gel free blot hands blotting reinvention is considered gold standard for protein detection and characterization although improvements to individual aspects of methodologies have been developed in recent years none has integrated entire process onto single platform proteinsimple assays sizing quantitative immunodetection as an alternative traditional ed vv analysis are performed on simon instrument that integrates automates esereser all manual operations associated with ights rights r rall introduction separated proteins immobilized capillary wall via most widely used accepted methodology proprietary photoactivated capture chemistry target then inc was first reported literature over identified primary antibody subsequent ica using horseradish peroxidase hrp conjugated secondary ago proven technique it plagued by poor reproducibility lack accurate quantitation chemiluminescent substrate molecular weight signal ameramer extensive ...