                                                                                                  APPLICATION NOTE 43325
         Analysis of pharmaceutical products for their 
         elemental impurities with the Thermo Scientific 
         iCAP RQ ICP-MS
         Authors                                  Introduction
         Julian Wills and Daniel Kutscher         Impurities in pharmaceutical products are of great concern not only due to 
         Thermo Fisher Scientific,                the inherent toxicity of certain contaminants, but also due to the adverse 
         Bremen, Germany                          effects that contaminants may have on drug stability and shelf-life. This 
                                                  necessitates the monitoring of organic and inorganic impurities throughout 
         Keywords                                 the pharmaceutical manufacturing process, from raw ingredients to final 
         FDA 21 CFR part 11,                      products. United States Pharmacopeia (USP) General Chapter <231>, 
         Microwave digestion,                     introduced in 1905, is a colorimetric test involving the co-precipitation of  
         Pharmaceutical compliance,               ten sulfide-forming elements and a visual color comparison to a  
         Pharmaceutical preparations,             10 ppm lead standard. The limitations of this test are well understood  
         United States pharmacopeia,              (non-specificity, the test is based on limited understanding of trace metal 
         USP 232, USP 233                         toxicity, etc.) so that consequently the USP published two new general 
                                                                                                st
                                                  chapters to replace <231> starting January 1 , 2018.
                                                                                        1
         Goal                                     • Chapter <232> Elemental Impurities  – Limits; defines the maximum limits of 
         To demonstrate the use of the              fifteen elements in pharmaceutical products
                          ™      ™                                                      2
         Thermo Scientific  iCAP  RQ              • Chapter <233> Elemental Impurities  – Procedures; defines how the testing 
         ICP-MS to accurately determine             for these elements should be performed.
         concentrations of elemental 
         impurities in pharmaceutical             From that date onward, all elemental impurity testing and all elemental 
         products brought into solution using     impurity testing must instead conform to the limits set out in Chapter <232>, 
         microwave digestion. All sample          using the procedures set out in Chapter <233>.
         preparation, measurement and data 
         evaluation to be compatible with the 
         guidelines defined in USP chapters 
         <232> Elemental Impurities – Limits 
         and <233> Elemental Impurities – 
         Procedures.
         In addition to the requirements described in the USP              All three drugs were brought into solution via a 
         documents, any analytical system used for the creation            microwave digestion procedure using an UltraWAVE 
         of analysis data for pharmaceuticals must also comply             closed vessel microwave digestion system (Milestone 
         with the US Food and Drug Administration’s (FDA) 21               Inc., Shelton, CT, USA). Different microwave recipes are 
         CFR Part 11 regulations regarding electronic records and          available to address specific sample matrices making 
         validation of electronic signatures. These regulations are        this the most universal method of sample preparation for 
         concerned with ensuring the integrity and authenticity            subsequent elemental analysis.
         of any electronic records and electronic signatures 
         that ‘persons create, modify, maintain, archive, retrieve         Samples of each drug (0.5 g) were weighed into 15 ml 
                      3
         or transmit’ . Control software used by analytical                disposable glass vials. For Drugs A and B, 3 ml of HNO3 
         instruments in pharmaceutical production must therefore           was added to each tube. For Drug C, 2 mL of HNO3 and  
         incorporate tools to maintain the integrity of the analytical     1 mL of H SO  was added to each vial. In compliance 
                                                                                     2    4
         method and subsequent results. In order to provide                with the repeatability requirements defined in USP 
         a transparent pathway to data generation, the control             <233>, six separate preparations of each material were 
         software should include support for audit trails and              prepared.
         electronic signatures as well as security features to 
         ensure that alterations cannot be made without clear              Sample vials were transferred into the microwave 
         indication of what has been changed, who changed it               digestion system which was then closed, pressurized 
         and why.                                                          with nitrogen at 40 bar and the temperature program 
                                                                           shown in Table 1 was launched. High pressure digestions 
         This note describes the effective application of the              are recommended due to the use of lower temperatures 
                            ™      ™
         Thermo Scientific  iCAP  RQ single quadrupole (SQ)                minimizing the loss of volatile elements.
         ICP-MS, to the detection and quantification of the  
         15 target elements specified in USP <232>, in                     Table 1. Closed vessel microwave temperature program used for
         accordance with the ICP-MS procedures described                   the dissolution of pharmaceutical products.
         in USP <233>. In order to generate data compliant with                 Step         Time (min)    Temperature      Power (W)
         the procedures described in 21 CFR Part 11, the                                                        (˚C)
                                    ™                                             1              15             200            1500
         Thermo Scientific Qtegra  Intelligent Scientific Data 
                  ™                                                               2              10             200            1500
         Solution  (ISDS) Software includes comprehensive 
         features for the pharmaceutical industry, such as 
         user access levels, audit trails, support for electronic          When sufficiently cooled, the clear, colorless digested 
         signatures as well as integrated, secure data                     material was transferred to polypropylene vials and 
         management.                                                       made up to 50 ml with ultrapure water. Each sample was 
         Sample preparation                                                then diluted by a factor of five into 15 ml polypropylene 
         It has been demonstrated that direct aqueous                      autosampler vials in a matrix of 1.2% HNO3 and 0.5%  
                                                                                           -1
         dissolution is suited for the preparation of water soluble        HCl + 200 μg·L  of gold to give a total dilution factor of 
         pharmaceutical samples before subsequent USP <233>                500 from the original solid sample. This diluent was used 
         compliant ICP-MS analysis. Indirect dissolution via closed        to ensure stability of the target elements in solution and 
         vessel microwave digestion, however, is recognized                efficient washout of these elements between samples 
         as the most universal sample preparation method for               from the sample introduction system.
         materials for subsequent elemental analysis by ICP-MS. 
         An important advantage of the closed vessel microwave             The samples were measured using an external calibration 
         approach is the retention of volatile elements, in particular     approach against calibration solutions prepared in 
         mercury that might otherwise be lost.                             the same diluent as the samples. The calibration 
                                                                           solutions contained all of the elements listed under the 
         Three pharmaceutical products were selected for                   Oral daily dose PDE (in µg·g-1) in USP <232>. Internal 
         analysis as part of this study:                                   standardization was applied, using Ga, In and Tl internal 
                                                                                                           -1
         Drug A: a phytotherapeutic (herbal) medicine                      standards at 5, 10 and 10 µg·L  respectively, added  
         Drug B: a vascular medicine                                       online via a T-piece.
         Drug C: an antianxiety medicine
           Calibration solution preparation                                                        As
           Sample analyses were carried out in accordance with 
           the requirements described in USP <233> Elemental 
           Impurities – Procedures. This document specifies that the 
           elements to be measured should be calibrated against 
           standard solutions at concentrations of blank, 0.5J and 
           2J where J = the concentration (w/w) of the element(s) 
           of interest at the target limit, appropriately diluted to the 
                                                  2
           working range of the instrument .
           Target limits for each of the USP <232> controlled 
           elements were calculated by dividing the permitted 
           daily exposure based on a 50 kg person (PDE) by the 
           maximum daily dose. For the three drugs used in this                                    Cd
           work, the maximum daily dose is 10 g.
           Table 2. Target limits (J) for the fourteen elements specified in  
           USP <232>.
            Element                   Oral daily dose PDE*        Target limit J 
                                                     -1                    -1
                                             (µg·day )               (µg·g )
            Cadmium                              5                     0.5
            Lead                                 5                     0.5
            Inorganic arsenic                   15                     1.5
            Inorganic mercury                   30                      3
            Iridium                             100                     10
            Osmium                              100                     10
            Palladium                           100                     10                         Hg
            Platinum                            100                     10
            Rhodium                             100                     10
            Ruthenium                           100                     10
            Chromium                           11000                  1100
            Molybdenum                         3000                    300
            Nickel                              200                     50
            Vanadium                            100                     20
            Copper                             3000                    300
           * PDE = permitted daily exposure based on a 50 kg person
           With this target limit taken into account, and as the 
           samples were diluted by a factor of 500 from the original                               Pb
           sample, two multielemental calibration solutions were 
           prepared at the concentration levels 0.5J and 2J in  
           2% HNO3.
           Results
           Calibration Curves
           Linear calibrations with low (sub ng·g-1) blanks were 
           obtained for all elements. Example calibration lines for 
           the ‘big four’ elements are shown in Figure 1.
                                                                                      Figure 1. Example calibrations for the ‘big four’ elements:  
                                                                                      As, Cd, Hg and Pb.
          Instrumental and Method Detection Limits                                    Sample analysis results
                              -1
          Single digit pg·g  instrumental detection limits (LoD)                      The final concentrations determined for each target 
          are typically obtained for all of the USP <232>                             element in the pharmaceutical products tested  
          defined elements (Table 3). Background equivalent                           (six repeat analyses per sample) are shown in  
          concentrations (BEC) for the 1.2% HNO3 and 0.5%                             Table 4. MDL and target limit (J) values are provided for 
          HCl calibration solution were also calculated. Low or                       comparison. Determined concentrations found to be  
          sub pg·g-1 detection limits (LOD) highlight the excellent                   less than the MDL are marked as ‘ required elements.                        from the six replicate analyses of the three drugs tested.
          However, while the instrumental detection limits in                                               Drug A  Drug B  Drug C          MDL      Target 
                                                                                       Element              (µg·g-1)  (µg·g-1)  (µg·g-1)   (µg·g-1)  Limit J 
          Table 3 illustrate the detection capabilities of the                                                                                       (µg·g-1)
          iCAP RQ ICP-MS for the analysis of the USP <232>                             Cadmium                defined elements.                 In each sample some elements were found to be below 
                             LOD           BEC          MDL        Target limit J     the calculated MDL but no element was found to be 
             Isotope        (ng·g-1)     (ng·g-1)      (µg·g-1)       (µg·g-1)        above the Target Limit, J.
            51
             V              0.0035        0.0629       0.014             10
            52Cr            0.007         0.042        0.008           1100           Drift
            60
             Ni             0.0012        0.0163       0.100             20           Following the requirement detailed in USP <233>, the 
            63
             Cu             0.0049        0.0910       0.186            300           read back concentrations for one of the calibration 
            75As            0.0009        0.0087       0.0005           1.5           standards analyzed before and after the sample solutions 
            95
             Mo             0.0026        0.0013       0.027            300           were compared. This comparison is made to ensure that 
            101
              Ru            0.0003        0.00005      0.025             10           the initial calibration remains valid over the entire analysis. 
            103Rh           0.0001        0.00005      0.026             10           The test is deemed to pass if the relative difference 
            105Pd           0.0036        0.0351       0.044             10           between two analyses of the calibration solution is less 
            111Cd           0.00001       0.00009      0.006            0.5           than 20%. All elements were found to be reproducible 
            189Os           0.0007        0.0003       0.043             10           over the complete analysis period (three hours in total) 
            193Ir           0.0005        0.0045       0.023             10           with relative standard deviation (RSD) between 0.1%  
            195Pt           0.0001        0.0002       0.024             10           to maximum 4%, and hence well within the USP <233> 
            202Hg           0.0099        0.0290       0.018             3            defined limit for the calibration solution containing a  
            208Pb           0.0009        0.0035       0.009            0.5           2J spike.
The words contained in this file might help you see if this file matches what you are looking for:

...Application note analysis of pharmaceutical products for their elemental impurities with the thermo scientific icap rq icp ms authors introduction julian wills and daniel kutscher in are great concern not only due to fisher inherent toxicity certain contaminants but also adverse bremen germany effects that may have on drug stability shelf life this necessitates monitoring organic inorganic throughout keywords manufacturing process from raw ingredients final fda cfr part united states pharmacopeia usp general chapter microwave digestion introduced is a colorimetric test involving co precipitation compliance ten sulfide forming elements visual color comparison preparations ppm lead standard limitations well understood non specificity based limited understanding trace metal etc so consequently published two new st chapters replace starting january goal limits defines maximum demonstrate use fifteen procedures how testing accurately determine these should be performed concentrations date o...