    Poster Reprint
    ASMS 2018
    TP-789
    Analysis of Amino Acids in 
    Animal Feed Matrices using the 
    Ultivo Triple Quadrupole LC/MS 
    System
    Jennifer Hitchcock1, Yanan Yang1, Gaëlle
    Bridon2                3
           , Hélène Lachance , Mathieu 
    D’Amours3
    1Agilent Technologies, Santa Clara, CA 
    2Agilent Technologies, Saint-Laurent, Quebec, 
    Canada,
    3Trouw Nutrition, Saint-Hyacinthe, Quebec, 
    Canada
                 Introduction                                                                                     Experimental
             Amino acids are called the building blocks of proteins                                          In this application, we demonstrate the sensitive and 
             and are necessary components of a balanced diet.                                                precise quantitation of 17 underivatized amino acids 
             For pets and farm animals, the diet must supplement                                             in an animal feed matrix using the novel Ultivo triple 
             the essential amino acids that include cysteine,                                                quad LC/MS. 
             histidine, isoleucine, leucine, lysine, methionine,                                             Reagent and Chemicals
             phenylalanine, threonine, tryptophan, tyrosine, and                                             All reagents used in this application were HPLC or 
             valine.  Pet food and animal feed nutrition companies                                           LCMS grade.  Acetonitrile was purchased from 
             therefore monitor their products for the presence of                                            Honeywell (Morristown, NJ, USA) and ultrapure water 
             these amino acids, along with the remaining non-                                                was sourced from a Milli-Q Integral system with an 
             essential ones, to ensure the quality of their food in                                          LC-Pak Polisher and a 0.22µm point-of-use membrane 
             providing a balanced diet.                                                                      filter cartridge (EMD Millipore, Billerica, MA, USA). 
                                                                                                             Formic acid and ammonium acetate were purchased 
             Historically, amino acids have been analyzed using a                                            from Fluka (Sigma-Aldrich Corp., St. Louis, MO, USA).  
             labor-intensive hydrolysis and derivatization based                                             Chemical standards were purchased from Pierce 
             sample preparation, which can limit the speed of                                                (ThermoScientfic, Waltham, MA, USA). 
             analysis.  However, it has been found that                                                      SamplePreparation
             underivatized amino acids can be analyzed with                                                  Animal feed was obtained from local suppliers and 
             excellent resolution and sensitivity using hydrophilic                                          ground up. A sample of 0.1 g was weighed in a 250 
             interaction chromatography (HILIC) with low pH                                                  mL Erlenmeyer flask and hydrolyzed using 25 mL 6 N 
             solvents and positive ion mode MS detection.                                                    HClwith phenol.  The flasks were placed under 
                                                                                                             nitrogen flow for 23 hours, then removed and allowed 
             The Ultivo is designed to address many of the                                                   to cool to room temperature.  The samples were 
             challenges faced by labs performing environmental                                               transferred to 50 mL volumetric flasks that had been 
             and food safety analyses.  Innovative technologies                                              rinsed with 0.1 N HCl prior to sample introduction.  
             housed within Ultivo created a reduced overall                                                  The flasks were filled to the line with 0.1 N HCl and 
             footprint while conserving the performance found in                                             mixed.  Aliquots were transferred to sample vials for 
             traditional systems (Figure 1).                                                                 introduction into the LC/TQ system.
                                              Innovations such as the Cyclone                                Data Analysis
                                              Ion Guide, Vortex Collision Cell,                              System control and data acquisition were performed 
                                              and the Hyperbolic Quads                                       by Agilent MassHunter Acquisition Software (C.01.00).  
                                              maximize quantitative                                          MS/MS transitions were obtained using MassHunter
                                              performance in a small package,                                Acquisition Optimizer software to determine optimal 
                                              enhancing instrument reliability,                              precursor and product ions, fragmentor voltages, and 
                                              robustness, and uptime.  Easy                                  collision energies upon injection of a neat solution of 
                                              access to the detector for                                     the compounds.  Data were analyzed using Agilent 
                                                                                                             MassHunterQuantitative Analysis Software (B.09.00) 
                                              replacement and the VacShield                                  and Qualitative Analysis Software (B.08.00).
                                              reduce the need for user                                       Method
                                              intervention for system                                        Table 1. The 1260 Infinity II Bio-Inert HPLC conditions.
                                              maintenance, making it 
                                              attractive to the non-expert MS                                 Column             Poroshell 120 HILIC-Z 2.1 x 150 mm, 2.7µm (p/n 6x3775-924)
                                              user to operate and maintain.                                   Column temp        25°C
                                              The MassHunter software suite                                   Injection volume   2 µL
                                              simplifies data acquisition, 
                                              method setup, data analysis and                                 Mobile phase       A: 20 mM ammonium acetate + 0.1% formic acid in water, pH 3
                                              reporting. This leads to faster                                                    B: 20 mM ammonium acetate in 90% acetonitrile, pH 3
                                              acquisition-to-reporting time,                                  Flow rate          0.50 mL/min
                                              increasing lab productivity and                                                    Time                  B%
                                              confidence in results.                                                             0                     100
                                                                                                              Gradient           11.5                  70
               Figure 1. Ultivo integrated into LC stack                                                                         12.0                  100
                                                                                                                                 15.0                  100 stop-time
                                                                                                                                 19.0                  100 post time
                                                                                                                                                                                                      2
             Experimental
         Capillary voltage on Agilent Jet Stream ESI source was set as 2000V with 0 nozzle voltage. The sheath gas temp is 400ºC 
         coupled with drying gas temp at 150ºC and 12 L/min. The nebulizer gas was set as 20 psi for the small underivatized 
         amino acids. Analysis was carried out with positive ionization and dynamic multiple reaction monitoring (dMRM) with 
         500 msecfor cycle time. Both Q1 and Q3 was set at unit resolution.  
         Table 2. Transitions for amino acid detection in dynamic MRM mode
                               Precursor    Product                 RT Window  Fragmen                                        Precursor    Product                      RT       Fragment
          Compound Name          (m/z)      (m/z)       RT (min)      (min)      tor(V)       CE (V)     Compound Name          (m/z)       (m/z)       RT (min)      Window     or(V)         CE (V)
                                                                                                                                                                       (min)
               Alanine            90.1        62.1        6.83           2           75          0            Leucine           132.1        44.1         4.62           2           85          14
               Alanine            90.1        44.1        6.83           2           75          0            Leucine           132.1        41.0         4.62           2           85          25
               Arginine          175.1       116.1       10.43           2          105          2            Leucine           132.1        30.0         4.62           2           85          4
               Arginine          175.1        70.1       10.43           2          105          8             Lysine           147.1       130.1        11.12           2           85          0
               Arginine          175.1        60.1       10.43           2          105          4             Lysine           147.1        84.1        11.12           2           85          6
             Aspartic Acid       134.1        88.0        9.03           2           75          0           Methionine         150.1       104.1         5.16           2           75          0
             Aspartic Acid       134.1        74.0        9.03           2           75          4           Methionine         150.1        61.0         5.16           2           75          14
             Aspartic Acid       134.1        70.0        9.03           2           75          6           Methionine         150.1        56.1         5.16           2           75          6
               Cystine           241.0       241.0       11.16           2          105          0           Methionine         150.1        28.0         5.16           2           75          26
               Cystine           241.0       152.0       11.16           2          105          0         Phenylalanine        166.1       120.1         4.23           2           85          4
               Cystine           241.0       120.0       11.16           2          105          0         Phenylalanine        166.1       103.1         4.23           2           85          22
               Cystine           241.0        74.1       11.16           2          105         25         Phenylalanine        166.1        91.1         4.23           2           85          32
            Glutamic Acid        148.1       148.1        8.27           2           85          0         Phenylalanine        166.1        77.0         4.23           2           85          36
            Glutamic Acid        148.1        84.0        8.27           2           85          6             Proline          116.1        70.1         6.01           2           85          6
            Glutamic Acid        148.1        56.1        8.27           2           85         22             Proline          116.1        43.1         6.01           2           85          25
            Glutamic Acid        148.1        41.0        8.27           2           85         18             Serine           103.1        88.1         7.63           2           65          0
               Glycine            76.0        48.0        7.36           2           65          0             Serine           103.1        60.1         7.63           2           65          0
               Glycine            76.0        30.0        7.36           2           65          0           Threonine          120.0        74.1         6.98           2           75          0
               Histidine         156.1       110.1        9.81           2           95          4           Threonine          120.0        56.1         6.98           2           75          6
               Histidine         156.1        95.1        9.81           2           95          6            Tyrosine          182.1       136.1         5.53           2           95          0
              Isoleucine         132.1        86.1        4.90           2           85          0            Tyrosine          182.1       119.1         5.53           2           95          10
              Isoleucine         132.1        44.1        4.90           2           85         16            Tyrosine          182.1        91.1         5.53           2           95          22
              Isoleucine         132.1        41.0        4.90           2           85         18            Tyrosine          182.1        77.0         5.53           2           95          34
              Isoleucine         132.1        30.0        4.90           2           85          6             Valine           118.1        72.1         5.84           2           75          0
               Leucine           132.1        86.1        4.62           2           85          0             Valine           118.1        55.1         5.84           2           75          14
             Results and Discussion
        Chromatography
        With the Poroshell HILIC-Z column and normal phase chromatography system , the 17 amino acids were well separated 
        within a 7-minute window (Figures 2 and 3).  Baseline separation of 0.3 minutes was also achieved for leucine and 
        isoleucine (Figure 4), and the retention times demonstrated excellent stability with RSDs of less than 0.5% for each 
        compound.
                                                                                                         Figure 3. Amino acids from Figure 2, with peak heights 
                                                                                                         normalized. 
                                                                                                                                                                                                      3
          Results and Discussion
      Linearity, Accuracy, and Reproducibility                              Figure 5 shows examples of calibration curves for 6 
                                                                            selected compounds, while 6 replicate injections of 3 
      The calibration concentrations ranged from 1.12 ppb to                selected compounds in matrix are shown in Figure 6, 
      18,843.76 ppb for the various analytes. Limits of                     demonstrating excellent precision.
      quantitation (LOQs), along with curve fit parameters, are 
                                               2 value greater than 
      given in Table 3.  Each curve had an R
      0.992 and responses showed excellent reproducibility 
      from run to run.  Calibration curve accuracies were within 
      11.5% of the expected concentration at the lowest level, 
      and they demonstrated RSDs within 20% at the LOQs and 
      within 5% at the higher levels. 
      Table 3. Calibration curve fit, LOQs, and signal-to-noise
      (S/N).
       Compound Name      Curve Fit      R2       LOQ      S/N at LOQ
                                                  (ppb)
           Alanine         Linear      0.9997     46.33       4.35
           Arginine      Quadratic     0.9920     8.71       10.75
         Aspartic Acid     Linear      0.9971     13.31       5.76
           Cystine         Linear      0.9969     61.04      13.51
         Glutamic Acid    Quadratic    0.9998     7.65        5.37
           Glycine         Linear      0.9986     75.07       3.51          Figure 5. Excellent precision demonstrated for 6 replicate 
           Histidine      Quadratic    0.9998     7.76       15.92
          Isoleucine       Linear      0.9961     1.36        5.88          injections of three amino acids in sample matrix. 
           Leucine         Linear      0.9978     1.36        4.03
            Lysine        Quadratic    0.9934     7.60       18.94              Conclusions
          Methionine       Linear      0.9977     1.55       14.57
         Phenylalanine     Linear      0.9974     1.72        4.82
            Proline        Linear      0.9954     1.15        6.03          Ultivo is an innovative triple quadrupole mass 
            Serine        Quadratic    0.9987     10.51       5.43          spectrometer that can minimize laboratory workspace 
           Threonine       Linear      0.9995     11.91       4.35          requirements and reduce maintenance challenges.  The 
           Tyrosine        Linear      0.9943     1.88        3.86          new LC/TQ system demonstrated the accurate and 
            Valine         Linear      0.9972     1.12        6.19
                                                                            sensitive detection of commonly monitored amino acids 
                                                                            in a pet food matrix, while the use of the Poroshell 120 
                                                                            HILIC-Z column enabled underivatized analytes to be 
                                                                            monitored, resulting in a simplified workflow.
                                                                                References
                                                                            • Agilent  Application  Note 5991-8582EN—Methods for the
                                                                               Analysis of Underivatized Amino Acids by LC/MS
                                                                            • Agilent Application Note 5991-8922EN—Quantitative Analysis of
                                                                               Underivatized Amino Acids in Plant Matrix by Hydrophilic
                                                                               Interaction Chromatography(HILIC)withLC/MSDetection
                                                                            • Agilent    Application    Note    5991-8816EN—Analysis       of
                                                                               Underivatized Amino Acids by LC/MS for Bioreactor Cell Culture
                                                                               Monitoring
                                                                            • Otter DE.   Standardised methods for amino acid analysis of
                                                                               food. British Journal of Nutrition (2012), 108, S230-S237.
                                                                            • Wu G. Dietary requirements of synthesizable amino acids by
                                                                               animals:  a paradigm shift in protein nutrition.   Journal of
      Figure 4. Calibration curves of selected compounds.                      AnimalScienceandBiotechnology(2014),5,34-45.
       This information is subject to change without notice.
       ©AgilentTechnologies,Inc.2018
The words contained in this file might help you see if this file matches what you are looking for:

...Poster reprint asms tp analysis of amino acids in animal feed matrices using the ultivo triple quadrupole lc ms system jennifer hitchcock yanan yang gaelle bridon helene lachance mathieu d amours agilent technologies santa clara ca saint laurent quebec canada trouw nutrition hyacinthe introduction experimental are called building blocks proteins this application we demonstrate sensitive and necessary components a balanced diet precise quantitation underivatized for pets farm animals must supplement an matrix novel essential that include cysteine quad histidine isoleucine leucine lysine methionine reagent chemicals phenylalanine threonine tryptophan tyrosine all reagents used were hplc or valine pet food companies lcms grade acetonitrile was purchased from therefore monitor their products presence honeywell morristown nj usa ultrapure water these along with remaining non sourced milli q integral ones to ensure quality pak polisher m point use membrane providing filter cartridge emd mill...