396x Filetype PPT File size 2.99 MB Source: users.ugent.be
Introduction
This presentation introduces gel permeation chromatography (GPC)
The mechanism of GPC shall be discussed
The method of analysing data shall be introduced
2
Types of Liquid Chromatography
Interactive adsorption, partition, ion exchange, etc
Non-interactive GPC, SEC, GFC
3
Nomenclature
Gel Permeation Chromatography GPC
Size Exclusion Chromatography SEC
Gel Filtration Chromatography GFC
4
What is a GPC System?
1. Reciprocating piston pump delivers eluent from
reservoir at a constant volumetric flow rate.
GPC is always an isocratic separation
2. Two position injection valve permits
introduction of sample solution without
interrupting solvent flow. GPC tends to use
larger injection volumes than HPLC (typically up
to 200ul)
3. GPC columns perform a separation based on
the molecular size of polymer molecules in
solution. Resolution and/or resolving range is
increased by use of multiple column systems
4. Detector responds to concentration of polymer
molecules eluting from the GPC columns
5
Additional Components Used in GPC
Concentration detectors
Differential refractometer (RI)
Ultraviolet absorbance (UV)
Evaporative light scattering or mass detector (ELS, EMD)
Infra-red (IR)
Molecular weight sensitive detectors
Viscometry
Light scattering
Additional systems
Online degasser
Autosampler
Column oven
Additional specific detectors
6